Home >
Heart Surgery Procedures >
Surgical Videos > Instructions for Use
Instructions
for Use
Product Indications for Use:
-
AXIUS™ VACUUM 2 Stabilizer Systems
- ACROBAT(tm) Mechanical Stabilizer System
- XPOSE™ Access Device 3
- VASOVIEW® 5 Endcoscpoic Vessel Harvesting System
AXIUS™ VACUUM 2 Stabilizer Systems
Indications for Use
Caution:
Federal (U.S.A.) law restricts this device to sale, distribution and use by
or on the order of a physician.
Indications:
Each Guidant Axius™ Vacuum 2 Stabilizer System is intended for use during
performance of cardiac surgical procedures through a sternotomy incision
approach. The AccessRail Platform in combination with an Activator Drive
Mechanism is used to spread the sternum, providing access and direct
visualization to the thoracic cavity. The AccessRail Platform also allows
for the organization of pericardial sutures. The Stabilizer isolates and
provides local immobilization of the target vessel on the beating heart.
Contraindications:
Do not position the stabilizer foot over a coronary artery, newly infarcted
or aneurysmal heart tissue.
Warnings and Precautions:
1. Physicians should be properly trained to perform cardiac surgical
procedures with Guidant instruments prior to use.
2. As with all surgical retractors, care should be taken to use only as much
retraction as necessary to provide adequate access and visualization.
3)Perform the anastomosis only when the stabilizer foot is properly seated
on the epicardium and adequate stabilization of the surgical site is
achieved.
4. When removing the Axius™ Vacuum Stabilizer from the epicardium, care
should be taken not to disrupt the anastomotic site.
5. Take care not to adjust or move the AccessRail Platform while pericardial
sutures are engaged in the platform.
6. Suture holder features are designed for "0" size sutures.
7.The Guidant Activator Drive Mechanisms are supplied non-sterile and must
be cleaned and sterilized prior to each use. Never use saline solution for
cleaning surgical instruments.
8. Steam sterilization is recommended to sterilize the Activator Drive
Mechanisms. Other methods of sterilization have not been demonstrated as
effective and are not recommended.
Refer to the instructions for use for complete listing of indications,
contraindications, warnings and precautions.
back to top
ACROBAT(tm) Mechanical Stabilizer
System
Indications for Use
Indications:
The Guidant ACROBAT(tm) Mechanical 2 Stabilizer System is intended for use
during performance of minimally invasive cardiac surgery through a
sternotomy incision approach. The AccessRail Platform in combination with an
Activator Drive Mechanism is used to spread the sternum, providing access
and direct visualization to the thoracic cavity. The AccessRail Platform
also allows for the organization of pericardial sutures. The ACROBAT(tm)
Mechanical 2 Stabilizer System isolates and provides local immobilization of
the target vessel on the beating heart.
Contraindications:
Do not position the stabilizer foot over a coronary artery, newly infarcted
or aneurysmal heart tissue.
Warnings and Precautions:
Physicians should be properly trained to perform cardiac surgical procedures
with Guidant instruments prior to use.
As with all surgical retractors, care should be taken to use only as much
retraction as necessary to provide adequate access and visualization.
Perform the anastomosis only when the stabilizer foot is properly seated on
the epicardium and adequate stabilization of the surgical site is achieved.
When removing the ACROBAT(tm) Mechanical 2 Stabilizer from the epicardium,
care should be taken not to disrupt the anastomotic site.
Take care not to adjust or move the AccessRail Platform while pericardial
sutures are engaged in the platform.
Suture holder features are designed for "0" size sutures.
The Guidant Activator Drive Mechanism is supplied non-sterile and must be
cleaned and sterilized prior to each use. Never use saline solution for
cleaning surgical instruments.
Steam sterilization is recommended to sterilize the Activator Drive
Mechanisms. Other methods of sterilization have not been demonstrated as
effective and are not recommended. Care should be taken to ensure that mount
and mount lever are clear of tissue when stabilizer is positioned on
retractor.
back to top
XPOSE™ Access Device 3
Indications for Use
Caution:
Federal (U.S.A.) law restricts this device to sale, distribution and use by
or on the order of a physician.
Indications:
The Guidant Axius™ Xpose™ Device 3 is intended to expose coronary arteries
during minimally invasive, off pump cardiac surgery through a sternotomy
incision approach with a Guidant Stabilization System.
Contraindications:
Do not engage the Guidant Axius™ Xpose™ Device 3 over a coronary artery,
newly infarcted or aneurysmal tissue. If hemodynamic instability is
experienced, gently return the heart to its resting position.
Indications:
The Guidant Xpose™ Access Device is intended to expose coronary arteries
during minimally invasive, off pump cardiac surgery through a sternotomy
incision approach with a Guidant Stabilization System.
Contraindications:
Do not engage the Guidant Xpose™ Access Device over a coronary artery, newly
infarcted or aneurysmal tissue. If hemodynamic instability is experienced,
gently return the heart to its resting position.
Warnings and Precautions:
Physicians should be properly trained to perform cardiac surgical procedures
with Guidant instruments prior to use. Many variables, including patient
anatomy, pathology, and surgical techniques, may influence procedural
outcomes. Patient and procedure selection is a responsibility of the medical
professional. The device is intended solely for use with Guidant
Stabilization Systems.
Refer to the instructions for use for complete listing of indications,
contraindications, warnings and precautions.
back to top
VASOVIEW® 5 Endcoscpoic Vessel Harvesting System
Indications for Use
VASOVIEW® 5 Harvesting Cannula Single-Use
Indications, Contraindications, Warnings, Precautions, Adverse Events
Caution: Federal (USA) law restricts this device to sale by or on the order
of a physician.
Indications:
The VASOVIEW 5 Harvesting Cannula has applications in minimally invasive
surgery and is primarily indicated for patients undergoing endoscopic
surgery for vessel harvesting. It is indicated for cutting tissue and
controlling bleeding through coagulation in general and cardiothoracic
surgery including minimally invasive direct coronary artery bypass (MIDCAB),
lower extremity and thoracoscopic procedures. Lower extremity procedures
include tissue dissection/vessel harvesting along the saphenous vein for use
in coronary artery bypass grafting and peripheral artery bypass.
Thoracoscopic procedures include exposure and dissection of structures
external to the parietal pleura, including nerves, blood vessels and other
tissues of the chest wall.
Contraindications:
The device is contraindicated in situations where minimally invasive surgery
is contraindicated. The Harvesting Cannula is not intended for contraceptive
coagulation of the fallopian tube but may be used to achieve hemostasis
following transection of the tube.
Warnings and Precautions:
1. Read all instructions carefully. Failure to properly follow the
instructions, warnings and cautions may lead to serious surgical
consequences or serious injury to the patient.
2. Minimally invasive surgical procedures should be performed only by
individuals adequately trained and familiar with such surgical techniques.
Consult medical literature regarding techniques, complications, and hazards
prior to performance of these procedures.
3. Sterility: The product is sterile unless the package is opened or
damaged. The product is designed for single use. Do not reuse or resterilize.
The method of sterilization is irradiation.
4. Before endoscopic instruments and accessories from different
manufacturers are employed in a procedure, verify compatibility and ensure
that electrical isolation and grounding of these instruments is not
compromised.
5. A thorough understanding of the principles and techniques involved in
electrosurgical procedures is essential to avoid shock and burn hazards to
both the patient and operator(s) and damage to medical instrumentation.
6. For use with the bipolar outputs of electrosurgical generators only!
7. Do not exceed 35 watts in any mode.
8. Use generators listed in the required special mode setting only.
9. Do not use with Codman Mallis generators.
10. All exposed metal components at the distal end of the scissors may
coagulate tissue. Ensure all exposed metal is within the field of vision and
contacting the tissue intended for coagulation during the application of
electrosurgical energy.
11. Handle the Endoscope carefully to avoid breakage. Regularly check the
orientation of the camera before advancing.
12. Advance the cannula gently to avoid damage to delicate tissue.
13. Always advance the C-ring and bipolar scissors under endoscopic
visualization. Ensure adequate visualization of bipolar scissor tips and
surgical site prior to application of electrosurgical energy.
14. Always inspect the surgical site for hemostasis. If hemostasis is not
present, appropriate techniques should be applied to achieve hemostasis.
15. In endoscopic procedures which use gas insufflation, venous gas embolism
is a very rare (approximately 1 in 10,000 cases) but potentially serious
complication that may occur. Its occurrence is signaled by cardiovascular
collapse (sudden, severe hypotension) and a precordial murmur. If gas
embolism is suspected during a procedure, discontinue gas insufflation and
place the patient in a left lateral and a slight Trendelenburg position.
back to top
7 mm Extended Length Endoscope and Dissection Tip
The 7 mm Extended Length Endoscope with Dissection Tip is indicated for
visualization of a surgical cavity and dissection in endoscopic procedures
and other minimally invasive surgical procedures allowing access for vessel
harvesting, and is primarily indicated for patients undergoing endoscopic
vessel harvesting for arterial bypass. It is indicated for patients
requiring endoscopic tissue separation of the extraperitoneal or
subcutaneous extremity and thoracic space. Extremity procedures include
tissue dissection along the saphenous vein and the femoral vessels.
Thoracoscopic procedures include exposure and dissection of structures
external to the parietal pleura, including nerves, blood vessels and other
tissues of the chest wall.
Contraindications:
The device is contraindicated in situations where endoscopic techniques and
minimally invasive surgery are contraindicated.
Warnings and Precautions:
1. Read all instructions carefully. Failure to properly follow the
instructions, warnings and cautions may lead to serious surgical
consequences or serious injury to the patient.
2. Endoscopic procedures should be performed only by physicians or surgical
professionals having adequate training and familiarity with endoscopic
techniques. Consult medical literature regarding techniques, complications,
and hazards prior to performance of these procedures.
3. The Endoscope is a reusable instrument that is supplied non-sterile.
Thoroughly clean and sterilize the Endoscope prior to each use. Follow
recommended cleaning and sterilization instructions as described in these
Instructions for Use.
4. The Endoscope is a fragile instrument. Handle carefully to avoid
breakage; bending of the shaft or dropping the Endoscope may damage the
optics or other internal components, making the Endoscope inoperable.
Sterilize and store in a protective tray whenever possible.
5. The Dissection Tip is sterile unless the package is opened or damaged.
The Tip is designed for single use. Do not reuse or resterilize the
Dissection Tip.
6. Whenever using endoscopic illumination equipment, ensure that the light
cable connectors never rest on flammable materials such as surgical drapes,
towels, etc.
7. Always inspect the surgical site for hemostasis. If hemostasis is not
present, appropriate techniques should be applied to achieve hemostasis.
8. In endoscopic procedures which use gas insufflation, venous gas embolism
is a very rare but potentially serious complication that may occur (in
approximately 1 in 10,000 cases). Its occurrence is signaled by
cardiovascular collapse (sudden, severe hypotension), and a precordial
murmur. If gas embolism is suspected during a procedure, discontinue gas
insufflation and place the patient in a left lateral and a slight
Trendelenburg position.
9. If Prevacuum or Gravity steam sterilization (autoclave) are used, the
Endoscope may have a shorter life due to harsher sterilization environment.
Inspect the Endoscope after each steam sterilization cycle for damage.
10. Do not cool a hot Endoscope after sterilization by rapidly exposing it
to air or liquid. Sudden temperature changes may cause glass components to
crack.
11. Do not "flash" steam sterilize (autoclave) the Endoscope. Flash (i.e.,
unwrapped) steam sterilization cycles introduce sudden temperature changes,
which may cause glass components to crack.
12. Sterilization with liquid chemical sterilants (e.g., STERIS) may not
convey the same sterility assurance as other sterilization methods.
back to top
VASOVIEW Short Port Blunt Tip Trocar (BTT)
Important: These instructions for use are designed to assist in the use of
the Short Port Blunt Tip Trocar (BTT). They are not a reference to
endoscopic surgery or techniques. Indications: This product has applications
for surgery in the leg for establishment of a port of entry for endoscopic
instruments.
Indications:
This product has applications for surgery in the leg for establishment of a
port of entry for endoscopic instruments.
Contraindications:
The device is contraindicated in situations in which minimally invasive
surgery is contraindicated.
Warnings and Precautions:
1. Read all instructions carefully. Failure to properly follow the
instructions, warnings, and cautions may lead to serious surgical
consequences.
2. Minimally invasive surgical procedures should be performed only by
persons having adequate training and familiarity with such surgical
techniques. Consult medical literature regarding techniques, complications,
and hazards prior to performance of these procedures.
3. Sterility: The product is sterile unless the package is opened or
damaged. The product is designed for single use. Do not reuse or resterilize.
The primary method of sterilization is irradiation. The alternate method is
ethylene oxide. For actual method used, see device label.
4. Before endoscopic instruments and accessories from different
manufacturers are employed in a procedure, verify compatibility and ensure
that electrical isolation and grounding of these instruments is not
compromised.
5. A thorough understanding of the principles and techniques involved in
electrosurgical procedures is essential to avoid shock and burn hazards to
both the patient and operator(s) and damage to medical instrumentation.
6. Balloon products must be treated with care. Damage to balloons by
instruments used during insertion and in the course of a procedure may
result in balloon rupture.
7. The balloon on the Short Port Blunt Tip Trocar contains a fully coated
natural rubber latex and may affect patients with latex sensitivities if the
coating becomes damaged.
8. Overinflation of the BTT Port balloon may result in balloon rupture. Do
not inflate more than 25 cc.
9. In endoscopic procedures which use gas insufflation, venous gas embolism
is a very rare (approximately 1 in 10,000 cases) but potentially serious
complication that may occur. Its occurrence is signaled by cardiovascular
collapse (sudden, severe hypotension), and a precordial murmur. If gas
embolism is suspected during a procedure, discontinue gas insufflation and
place the patient in a left lateral and a slight Trendelenburg position.
Refer to the instructions for use for complete listing of indications,
contraindications, warnings, and precautions. |